EasyManua.ls Logo

Impulse Dymanics OPTIMIZER Smart System - EMERGENCY PROGRAMMING; Emergency Programming When the Programmer is OFF

Impulse Dymanics OPTIMIZER Smart System
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61
19. EMERGENCY PROGRAMMING
In an emergency, the OMNI Smart Programmer can program the OPTIMIZER Smart IPG with a
safe parameter set [Standby (OOO) mode, CCM OFF]. This Emergency Programming can be
performed even if the Programmer is turned OFF (either the Tablet PC is OFF or is not
functioning).
19.1 Emergency Programming When the Programmer is OFF
Note: Even if the Programmer is OFF, the Programmer Wand must still be plugged into
the OMNI II Programmer Interface Box before using the Emergency Programming
function.
To program the OPTIMIZER Smart IPG with a safe parameter set when the OMNI Smart
Programmer is OFF:
Place (or replace, if necessary) the Programmer Wand over the OPTIMIZER
Smart IPG implant site.
Press the Emergency Programming button on the Programmer Wand.
If the Emergency Programming is successful, the Emergency Programming Indicator
Light on the Programmer Wand will flash for a few seconds.
Warning: When the Emergency Programming function is used with the OMNI Smart
Programmer OFF, the Programmer Wand uses a long-life lithium battery
inside the OMNI II Programmer Interface Box. The charge of this battery is
consumed only if the Programmer Interface is unplugged from the Tablet PC
or if the OMNI Smart Programmer is OFF when performing Emergency
Programming. Though the estimated average longevity for the lithium battery
inside the OMNI II Programmer Interface Box is 54 months, the voltage of the
lithium battery inside the OMNI II Programmer Interface Box should be
periodically checked to ensure that the battery is not depleted. The voltage of
this battery can be read by selecting the Interface battery button under the
Settings tab of the Parameters Bar on the OMNI Smart Programmer software
screen. If the battery voltage reading is less than 2.5 V, the battery inside the
OMNI II Programmer Interface Box must be replaced. If battery replacement
is required, please contact your Impulse Dynamics representative. This battery
is not intended to be replaced by the operator.
To perform any other operation, it is necessary to turn ON the OMNI Smart Programmer
System.