58
15.2.1 Cancel
If any parameter values have been modified but not yet programmed into the
OPTIMIZER Smart IPG, the Cancel command will reset the parameter values to
the last interrogated/programmed set.
To cancel modifications, perform one of the following actions:
Select the Cancel button on the Programming Bar. This button is
enabled only if a parameter value has been modified, or
Select the Cancel button on the Tool Bar, or
Press the keyboard shortcut <Esc>
Please note that after executing a Cancel command, the parameter values in the
OMNI Smart Programmer Software screens become black, since they are the
same as the parameter values in the OPTIMIZER Smart IPG. If no device was
interrogated and parameter data were loaded from a .tip file, the displayed
parameter values are the values stored in the standard (.tip file).
15.2.2 Undo
If the OPTIMIZER Smart IPG has been reprogrammed with a new set of
parameter values, the Undo button resets the parameter values to the prior set of
programmed values.
To undo the most recent programming, perform one of the following actions:
Select the Undo button on the Programming Bar. This button is enabled
only after the execution of a Program command, or
Select the Undo button on the Tool Bar. This button is enabled only after
the execution of a Program command, or
Press the keyboard shortcut <Ctrl+U>
If the Undo operation is successful, the OMNI Smart Programmer will display the
messageProgramming OK”.
However, if the Programmer Wand is not well positioned over the implant site,
the Undo operation may fail. If a failure in communication occurs, the
Programmer will display the message “Programming Error” along with the
options Retry and Cancel.
Retry can be performed by repositioning the Programmer Wand and selecting the
Retry button.
16. USING STANDARDS
Some standard parameter combinations are useful in certain clinical situations. Specific
programs can be stored as standards (also known as user presets). Files containing a standard
have a special format that the OMNI Smart Programmer Software can interpret. The file
extension used for them is “.tip”. The Open and Save commands of the OMNI Smart
Programmer read and write data from and to .tip files. As such, the OMNI Smart Programmer
Software can also be used as an editor of standards.
This section includes information on the use and storage of parameter value configurations.

Table of Contents