Current Consumption
1
Current consumption of the OPTIMIZER Smart IPG is strongly dependent on the energy delivered by the CCM pulse train.
Safe Mode
When the device encounters conditions considered to be the
result of faulty device circuitry or software, it will switch to
DOWN mode. In DOWN mode, the device is completely
quiescent; CCM signals are not delivered and cardiac events are
not sensed. To take the device out of this mode, a
device reset
must be performed under physician supervision.
Programmable Parameters
Standby mode: no events are sensed and no CCM impulse trains
are delivered
Active mode where the device senses atrial, ventricular and
Local Sense events and is capable of CCM signal delivery
Active mode where the device senses ventricular and local sense
events and is capable of CCM signal delivery
without the need
for the detection of atrial sense events.
Atrium sense amplifier sensitivity
13 possible between 0.1 mV to 5.0 mV
1
Ventricle sense amplifier
sensitivity
18 possible between 0.1 mV to 10.0 mV
Ventricle sensing polarity
Atrium refractory period
1
Between 148 ms to 453 ms in 8 ms increments
Ventricle refractory period
Between 148 ms to 453 ms in 8 ms increments
1
: Active only when the OPTIMIZER Smart IPG is in Active ODO-LS-CCM mode.