12.3.2 Graph View
A graphical view is available on the OMNI Smart Programmer software to
display some of the OPTIMIZER Smart IPG parameters in an easy-to-understand
graph.
This window displays the following parameter values:
• Right Heart Sensing
o Atrial Sensitivity
o Ventricular Sensitivity
o Atrial Refractory Period (PVARP)
o Ventricular Refractory Period
o Short AV Limit
o Long AV Limit
• Local Sense
o LS Sensitivity
o LS Alert Window Start
o LS Alert Window Width
o Pre Atrial LS Refractory Period
o Post Atrial LS Refractory Period
o Pre Ventricular LS Refractory Period
o Post Ventricular LS Refractory Period
o Post LS Refractory Period
• CCM Signal
o Number of pulses in CCM pulse train
o Trigger-to-CCM signal delay (coupling interval)
o CCM signal initial amplitude
o Phase duration of individual phases in each CCM pulse in CCM pulse
train
This window also displays a marker at the top left of the display indicating cursor
position in msec. It also places the text ‘Modified’ alongside the marker when a