11.1.2 OMNI II
The buttons contained within the OMNI II box are for use with the
OPTIMIZER IVs IPG. For details please see OPTIMIZER IVs Physician’s
Manual.
• Clinical Mode: For regular programming of the OPTIMIZER IVs IPG
• Remote Mode: For remote control of another OMNI Smart Programmer
using the OMNI II Programmer software.
• Remote Listener: For remote watching of another OMNI Smart
Programmer using the OMNI II Programmer software.
11.1.3 OMNI Smart
The buttons contained within the OMNI Smart box are for use with the
OPTIMIZER Smart IPG.
• 3-Lead Clinical Mode: For regular programming of the OPTIMIZER
Smart IPG (OVO-LS-CCM Mode is disabled)
• 3-Lead Remote Mode: For remote control of another OMNI Smart
Programmer using the OMNI Smart Programmer software in 3-Lead
Clinical Mode (see Section 31).
• 2-Lead Clinical Mode: For regular programming of the OPTIMIZER
Smart IPG
• 2-Lead Remote Mode: For remote control of another OMNI Smart
Programmer using the OMNI Smart Programmer software 2-Lead Clinical
Mode (see Section 31).
• Remote Listener: For remote watching of another OMNI Smart
Programmer using the OMNI Smart Programmer software (see Section
31).
11.1.4 Miscellaneous
The buttons displayed at the bottom of the Selection screen are for miscellaneous
system functions.
• System Admin: This selection allows special settings of OMNI Smart
Programmer. It is to be used by Impulse Dynamics technical staff only and
is not needed for regular clinical use.
• Configuration: Selection of the Configuration button opens a pop-up
window for additional settings.
o Set Bluetooth: Setup for Bluetooth settings needed for printer
connection.
o Network Configuration: Setup for network settings (LAN and
WLAN) needed for remote control and remote mode of OMNI Smart
Programmer.
o Set Default Printer: Selecting the default printer to be used by the
OMNI Smart Programmer.