60
In all cases, an Open window will first pop up where the name of the marker file
associated with the device will need to be selected.
Then, a Save window will appear which will allow the user to enter the name and, if
desired, create a new folder for the file that is to be exported.
18. LOG AND ECG RECORDING
18.1 Log
The OMNI Smart Programmer keeps a log of all the interactions that take place between
it and an OPTIMIZER Smart IPG. This record can be used as an index to provide fast
access to specific data associated with the communications. Bookmarks can also be set in
order to have fast access to specific events and conditions that deserve special attention in
the patient’s ECG.
The following is a description of the basic features of the log:
The log file for a particular OPTIMIZER Smart IPG is created when the device is
interrogated for the first time.
Each communications event that takes place between the OMNI Smart
Programmer and the OPTIMIZER Smart IPG appears in the Log Bar window,
together with the date and time of each interaction.
More information regarding an event in the log may be obtained by double-
clicking on the event in the Log Bar. If there are data associated with the selected
event, the following occurs:
o If an Interrogation or Programming event is chosen, the parameter values
displayed by the OMNI Smart Programmer are the set of values present at the
moment the selected event occurred.
o If an event involving Statistics (clear/load/reset) is chosen, the Statistics Bar
values are set to the values present at the moment the selected event occurred.
For all other recorded events (begin/end of Marker Mode, etc.), a message
appears indicating that the event has no more associated data to display.
18.2 ECG Recording
To record a patient’s ECG:
Select the Log Files button on the Task Bar
Select the Recording button on the Tool Bar
The same procedure must be used to stop recording the ECG.
When the OPTIMIZER Smart IPG is set to Marker Mode, the ECG starts recording
automatically. When Marker Mode is stopped, the ECG stops recording.
In the recorded ECG, Markers are displayed in the same way as in the Marker Window.
(See Section 23 for more information on Markers.)
The recorded ECG segments are displayed in the Marker Window.