10.11 Routine Cleaning
Warning: DO NOT attempt to sterilize an OMNI Smart Programmer or the Programmer
Wand because any such attempt could severely damage the equipment.
Warning: DO NOT submerge any part of the OMNI Smart Programmer in water.
Damage to the unit may result. The OMNI Smart Programmer System is not
protected against ingress of water or humidity (ingress protection rating
IPX0).
Warning: Always turn the OMNI Smart Programmer off prior to cleaning.
Following each use, it is recommended that a soft cloth dampened with a germicidal
cleaning solution be used to wipe the exterior case of the programmer, the ECG cables
and, particularly the Programmer Wand. Do not use solvents or cleaning cloths
impregnated with chemical cleaning agents.
10.12 Storage and Handling
The OMNI Smart Programmer and Programmer Wand are designed to function normally
after they have been exposed (while being packaged for transport) to the following
environmental extremes: (1) -4°F to +158°F, (2) relative humidity 10% to 100% (with or
without condensation), (3) atmospheric pressure 500 hPa to 1060 hPa.
Recommended conditions for normal use are: (1) 32°F to +131°F, (2) relative humidity
between 20% and 75%, and (3) atmospheric pressure between 700 hPa and 1060 hPa.
10.13 Additional Notes
The OMNI Smart Programmer has an isolated patient connection with defibrillation
protection.
In general, contact between the Programmer Wand and the patient’s skin should be
avoided due to the risk of cross-contamination.
OMNI II Programmer Interface Box has a battery for back-up safe mode programming.
This battery is not intended to be replaced by the operator. It can only be replaced by
designated Impulse Dynamics service personnel.
Note: The OMNI Smart Programmer System does not contain any user serviceable
parts. It must be replaced if it fails to operate in accordance with its specifications.
Warning: DO NOT discard the OMNI II Programmer Interface Box or Programmer
Wand in the trash. The OMNI II Programmer Interface Box contains a
Lithium battery as well as non-RoHS components. The Programmer Wand
contains non-RoHS components. If disposal of the OMNI II Programmer
Interface Box or Programmer Wand is necessary, properly dispose of the
OMNI II Programmer Interface Box or Programmer Wand in accordance with
local regulations governing the disposal of such material.
The OMNI Smart Programmer is classified as Class II equipment when connected to the
supply mains, Internally Powered ME equipment when laptop power supply not
connected to mains.
The Programmer Wand is classified as a Type BF applied part and ECG channel is
classified as a defibrillation-proof Type BF applied part.