21. RESETTING THE OPTIMIZER SMART IPG
The OPTIMIZER Smart IPG has protective mechanisms that maintain the internal consistency of
the system. These mechanisms detect when an internal discrepancy (for instance, clocks not
oscillating at the expected frequency) occurs.
If in the unlikely event a malfunction of this type occurs, the OPTIMIZER Smart IPG will place
itself into a safe state referred to as “DOWNmode. In “DOWN” mode, the OPTIMIZER Smart
IPG does not deliver CCM signals, and in some cases it does not even sense cardiac events. This
status can be changed only by resetting the OPTIMIZER Smart IPG using the OMNI Smart
Programmer under physician supervision.
To reset the OPTIMIZER Smart IPG:
Place (or replace, if necessary) the Programmer Wand over the OPTIMIZER Smart IPG
implant site.
Select the Reset button on the Tool Bar.
If the OPTIMIZER Smart IPG is successfully reset, the Programmer will display the message
Reset Device OK”.
However if the Programmer Wand is not well positioned over the implant site, the resetting
operation may fail. If a failure in communication occurs, the Programmer will display the
message “Reset Device Erroralong with and the options Retry and Cancel.
Retry can be performed by repositioning the Programmer Wand and selecting the Retry button.
IMPORTANT: If a unit is found in “DOWN” mode, please document the event before resetting
the device. After recording the cause of the reversion displayed by the OMNI Smart
Programmer, contact your Impulse Dynamics representative. Also, please provide details of the
programmed mode in which the reversion occurred, and any conditions that may have caused the
device to revert to “DOWN” mode.
22. INITIATING CCM DELIVERY
22.1 Operating Mode Options
The operational status of the OPTIMIZER Smart IPG is set by the Mode parameter on
the Device Bar. Selecting the Mode parameter will cause the Mode pop-up menu to
appear.