25.2 Parameter Modifications in Active OVO-LS-CCM Mode
Below is a list of parameters that are modified when the OPTIMIZER Smart IPG is set to
Active OVO-LS-CCM:
The display of PVC, AT, Long and Short AV flags in Marker Mode are disabled
CCM Inhibit on PVC is disabled
CCM Inhibit on Long AV is disabled
CCM Inhibit on Short AV is disabled
CCM Inhibit on Atrial Tachycardia is disabled
CCM Inhibit on Ventricular Tachycardia is enabled, with a programmable range,
but a maximum of 110 bpm
The minimum right ventricle (V) sensitivity is limited to 1.0 mV
The maximum CCM delay is limited to 45 ms
The minimum LS sensitivity is limited to 1.0 mV
The maximum LS Alert Window Width is limited to 30 ms