Figure 44: Overview Tab in Active OVO-LS-CCM Mode
Figure 45: A/V Tab in Active OVO-LS-CCM Mode
25.2 Parameter Modifications in Active OVO-LS-CCM Mode
Below is a list of parameters that are modified when the OPTIMIZER Smart IPG is set to
Active OVO-LS-CCM:
• The display of PVC, AT, Long and Short AV flags in Marker Mode are disabled
• CCM Inhibit on PVC is disabled
• CCM Inhibit on Long AV is disabled
• CCM Inhibit on Short AV is disabled
• CCM Inhibit on Atrial Tachycardia is disabled
• CCM Inhibit on Ventricular Tachycardia is enabled, with a programmable range,
but a maximum of 110 bpm
• The minimum right ventricle (V) sensitivity is limited to 1.0 mV
• The maximum CCM delay is limited to 45 ms
• The minimum LS sensitivity is limited to 1.0 mV
• The maximum LS Alert Window Width is limited to 30 ms