12.5 Main Tool Bar
The Main Tool Bar is displayed under the ECG Window. It contains the buttons of the
following OMNI Smart Programmer commands (Whenever a shortcut alternative exists
for a command, it is denoted next to the command name in parentheses):
Interrogate (Ctrl+I): Reads the current parameter values of the OPTIMIZER
Smart IPG. These values become the parameter values of the OMNI Smart
Programmer.
Program (Ctrl+P): Sets the parameter values of the OPTIMIZER Smart IPG
with the current parameter values of the OMNI Smart Programmer. This
command is enabled only if no parameter conflict occurs.
Cancel (Esc): When any parameter value of the Programmer is modified, the
Cancel command becomes available. Selecting this command will reset the
parameter values to the last interrogated/programmed ones. If no device was
interrogated and data were loaded from a .tip file, the programmer parameters are
set to the values defined in the file.
Undo (Ctrl+U): If the OPTIMIZER Smart IPG has been reprogrammed with a
new set of parameter values, this command resets the OPTIMIZER Smart IPG to
the previously programmed set of values.
Restore: If the OPTIMIZER Smart IPG has been reprogrammed to a new mode,
this command restores the OPTIMIZER Smart IPG to the values of the session’s
first interrogation.
Reset: Resets the OPTIMIZER Smart IPG. This command should be used only if
the OMNI Smart Programmer reports that the OPTIMIZER Smart IPG being
interrogated is in the “DOWN” mode [Standby (OOO), no CCM reversion mode].
Marker (Ctrl+M): Sets the OPTIMIZER Smart IPG in Marker Mode (i.e.,
Marker Events are displayed in the ECG Window for every sensed and generated
event.
Open (Ctrl+O): Reads a standard (.tip) file containing saved parameter values
stored on the Programmer’s tablet PC. Opens a window allowing the user to
choose the name and location of the saved standard (.tip) file. Reads device data
from a .tip file.
Save: Writes a standard (.tip) file containing the current parameter values to the
Programmer’s tablet PC. A window pops up asking for the name and location of
the standard (.tip) file.
Time: Opens the Time window which shows the current time settings of the real-
time clocks in the OPTIMIZER Smart IPG and the OMNI II Programmer Tablet
PC. This window also allows the time of the IPG to be set with PC time. (The
time setting of the OPTIMIZER Smart IPG is used by the CCM signal delivery
scheduling mechanism to turn CCM signal delivery on and off on a daily basis).