13.2 Interrogating the OPTIMIZER Smart IPG
To read the parameter values of the OPTIMIZER Smart IPG:
Place (or replace, if necessary) the Programmer Wand over the OPTIMIZER
Smart IPG implant site.
Perform one of the following actions:
o Press the Interrogate button on the Programmer Wand, or
o Select the Interrogate button on the Tool Bar, or
o Select the Interrogate button on the Programming Bar, or
o Press the keyboard shortcut <Ctrl+I>
If the interrogation is successful, the OMNI Smart Programmer will display the message
Interrogation OK”. The device model, serial number, current CCM therapy status, and
current battery level of the OPTIMIZER Smart IPG being interrogated will appear on the
Device Bar of the Programmer’s screen.
However if the Programmer Wand is not well positioned over the implant site, the
interrogation operation may fail. If a failure in communication occurs, the Programmer
will display the messageInterrogation Error” along with the options Retry and
Cancel.
Retry can be performed by repositioning the Programmer Wand and selecting the Retry
button or by pressing the Interrogate button located on the Programmer Wand.
14. MODIFYING PARAMETER VALUES
Parameter values are found on the Parameters and Follow Up bars and can be viewed and
modified by selecting one of the tabs on these bars and then choosing one of the parameters on
the panel.
To modify a parameter value:
Select the tab where the parameter to be modified appears.
Select the parameter value to be modified. A window with all of its possible values will
appear.
Select the new value from the list. This value will become the new parameter value.
Checkboxes are used for enabling/disabling parameters. A check () symbol indicates
that the option is selected. In order to change the option, select the box to the left of the
parameter name.
If the setting of a parameter value is not applicable in the context of other selected parameters, its
value is not displayed.
Note: Parameter values displayed on the programmer screen are not transferred to the IPG until
the Program command is issued.
Note: It is not necessary to keep the Programmer Wand over the OPTIMIZER Smart IPG while
the parameters are being modified on the programmer screen. Once the modified parameter is
ready to be programmed into the device, ensure correct placement of the Programmer Wand over
the implant site and then issue the Program command.