69
Note: When the parameter Scheduled is set to a specific value, the OMNI Smart Programmer
software automatically calculates and sets the exact times for the On Time and Off time
parameters using the default parameters for the Start Time and End Time.
For instance, if CCM therapy rate is set to 7 hours per day distributed over 24 hours, it sets the
following standard scheduling parameters:
Start Time 0 h 0 m
End Time 23 h 59 m
On Time 1 h 0 m
Off Time 2 h 25 m
Scheduled 7 h
25. Active OVO-LS-CCM Mode
25.1 Deactivated Parameters in Active OVO-LS-CCM Mode
When the OPTIMIZER Smart IPG is set to Active OVO-LS-CCM mode, the device no
longer requires the detection of an atrial event in order to deliver CCM therapy. Thus the
parameters associated with atrium heart sensing and atrial events are deactivated.
Figure 42: Sensing Tab in Active OVO-LS-CCM Mode
Figure 43: A/V Setup Tab in Active OVO-LS-CCM Mode