EasyManua.ls Logo

Impulse Dymanics OPTIMIZER Smart System - Emergency Programming

Impulse Dymanics OPTIMIZER Smart System
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
108
Emergency Programming
PARAMETERS RELATED TO CONTROLLING RIGHT HEART SENSING
Mode OOO
Atrial Sense Amplifier Sensitivity
1.3 mV
Ventricular Sense Amplifier Sensitivity 2.0 mV
Ventricular Sensing Polarity Bipolar
Atrial Sensing Polarity Bipolar
Ventricular Refractory Period
250 ms
Post-Ventricular Atrial Refractory Period 250 ms
CCM PULSE TRAIN ACTIVATION
CCM Pulse train enable OFF
CCM PULSE TRAIN TIMING
Number of pulses 2
Train delay
35 ms
Phase 1 duration 5.14 ms
Phase 2 duration 5.14 ms
Phase 1 polarity Positive
Phase 2 polarity Negative
CCM Pulse Amplitude 7.5 V
CCM signal delivery channel
LS, V
Interval 0 ms
CCM INHIBIT ALGORITHM
Programmable parameters to inhibit CCM signal delivery
CCM Inhibit Count 2 beats
Short AV Delay 70 ms
Long AV Delay 398 ms
Atrial tachycardia rate 154 bpm

Table of Contents