15. PROGRAMMING
15.1 Programming the OPTIMIZER Smart IPG
Programming the OPTIMIZER Smart IPG with the modified parameter values is allowed
only if no parameter conflict occurs.
The Program button will indicate whether a modified parameter value is allowed in the
following manner:
Disabled, if there is a parameter conflict,
Blinking blue, if parameter values have been modified, and there is no parameter
conflict.
To program the modified parameter values:
Place (or replace, if necessary) the Programmer Wand over the OPTIMIZER
Smart IPG implant site.
Perform one of the following actions (these actions will be enabled only if there is
no parameter conflict):
o Press the Program button on the Programmer Wand or
o Select the Program button on the Tool Bar, or
o Select the Program button on the Programming Bar,
o Press the keyboard shortcut <Ctrl+P>
If the programming is successful, the OMNI Smart Programmer will display the message
Programming OK”.
However if the Programmer Wand is not well positioned over the implant site, the
programming operation may fail. If a failure in communication occurs, the Programmer
will display the message “Programming Error” along with the options Retry and
Cancel.
Retry can be performed by repositioning the Programmer Wand and selecting the Retry
button or by pressing the Program button located on the Programmer Wand.
Please note that after executing a Program command, the modified parameter values on
the OMNI Smart Programmer Software screen will become black, indicating that they are
now the programmed parameter values of the OPTIMIZER Smart IPG.
15.2 Cancel and Undo Functions
Modifications to the parameter values can be reset to the previous values in two different
ways. The method used to reset the modified parameters depends on whether or not the
modified parameters have been programmed into the OPTIMIZER Smart IPG.