73
On Inhibition
o Causes
AT: number of atrial tachycardia beats detected during scheduled
CCM delivery
PVC: number of PVCs detected during scheduled CCM delivery
Long AV: number of times Long AV condition detected during
scheduled CCM delivery
Short AV: number of times Short AV condition detected during
scheduled CCM delivery
LS: number of Local Sense events outside the LS Alert Window
during scheduled CCM delivery
LS Absence: number of Local Sense events not detected during
scheduled CCM delivery
VT: number of ventricular tachycardia beats detected during
scheduled CCM delivery
o Noise Episodes
A Noise: number of times atrial noise detected during scheduled
CCM delivery
V Noise: number of times ventricular noise detected during
scheduled CCM delivery
Off General: Displays the same list of statistical counters as detailed
under On General (except for Trains Delivered). Statistics for when
the OPTIMIZER Smart IPG is scheduled not to deliver CCM therapy
Off Inhibition: Displays the same list of statistical counters as detailed
under On Inhibition. Statistics for when the OPTIMIZER Smart IPG is
scheduled not to deliver CCM therapy
Last Session: Statistics for Last Delivery V, Last Delivery Trains,
Percentage, and Battery Discharge Episodes.
26.1.2 Statistics Buttons
Read: Reads statistics from IPG.
Reset: Resets statistics stored in IPG.
Numeric: Displays a table of the numerical statistics from the active tab.
Graphic: Displays a histogram of the statistics from the active tab.