34.6.6 Numerical Code 5
When Numerical Code 5 is displayed, it means that the temperature of the
OPTIMIZER Smart IPG at the beginning of a charging session is greater than
102.2°F. This Numerical Code may also be displayed if the temperature of the
OPTIMIZER Smart IPG during charging increases by more than 3° for over 10
minutes. If this Numerical Code is displayed repeatedly by the charger over the
course of several days, please contact your Impulse Dynamics representative.
34.6.7 Numerical Code 6
When Numerical Code 6 is displayed, it means that the OPTIMIZER Mini
Charger has detected a fault within the charger itself. If this Numerical Code is
displayed by the charger, please contact your Impulse Dynamics representative.
34.6.8 Numerical Code 7
When Numerical Code 7 is displayed, it means that the OPTIMIZER Mini
Charger has determined that it is attempting to be used on an unrecognized
device. If this Numerical Code is displayed by the charger, please verify the
charging wand is placed over the implanted OPTIMIZER Smart IPG. If this code
is still displayed after the charging wand is placed over the implanted
OPTIMIZER Smart IPG, please contact your Impulse Dynamics representative.
34.6.9 Numerical Code 8
When Numerical Code 8 is displayed, it means that the OPTIMIZER Mini
Charger has detected that even after attempting to charge a deeply discharged
battery in the OPTIMIZER Smart IPG, the level of the battery’s charge remains
low. If this Numerical Code is displayed by the charger, please contact your
Impulse Dynamics representative.
Note 1: Whenever Numerical Code 0 or 5-8 is displayed, the charger
automatically terminates the charging process.
Note 2: When Numerical Code 0 or 5 occurs, the charger can be set to a special
state to conduct a charging session. However, this option is only accessible to
Impulse Dynamics service technicians.
34.7 Cleaning
The OPTIMIZER Mini Charger should only be cleaned with disinfectant wipes as
needed.
Warning: DO NOT submerge any part of the OPTIMIZER Mini Charger in water.
Damage to the unit may result. The OPTIMIZER Mini Charger has limited
protection against ingress of water or humidity (ingress protection rating
IP22).
Warning: DO NOT sterilize any part of the OPTIMIZER Mini Charger because any
such attempt could severely damage the equipment.