25.3 Marker Events in Active OVO-LS-CCM Mode
With the OPTIMIZER Smart IPG is set to Active OVO-LS-CCM in Marker mode, atrial
events are ignored. As such, atrial event markers are not displayed in the ECG Window.
Figure 46: ECG Window in Marker Mode (Active OVO-LS-CCM Mode shown)
26. STATISTICS
The OPTIMIZER Smart IPG accrues a statistical record of events and conditions that occur as it
functions. This record can be loaded into the OMNI Smart Programmer, and appears in the
Statistics window of the OMNI Smart Programmer software, indicating the number of
occurrences of each event type.
Events considered are as follows:
• Atrial: the number of sensed atrial events.
• Ventricular: the number of sensed right-ventricular events.
• A Noise: the number of times atrial noise was detected.
• V Noise: the number of times ventricular noise was detected.
• Short AV: the number of times a Short AV condition was detected.
• Long AV: the number of times a Long AV condition was detected.
• PVC: the number of times a PVC was detected.
• AT: the number times the instantaneous atrial rate exceeded the Atrial Tachycardia rate
limit.
• VT: the number times the instantaneous ventricular rate exceeded the Ventricular
Tachycardia rate limit.
• Trains Delivered: the number of delivered CCM signals.
• LS Inhibit: the number of CCM signal delivery inhibitions due to Local Sense events
detected outside the LS Alert Window.
• LS Absence: the number of CCM signal delivery inhibitions because Local Sense events
were not detected.
• LS in Alert: the number of Local Sense events inside the LS Alert Window.