Put On the O-Rings
9-6 CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide
Put On the O-Rings
1. Locate the o-rings for the handpiece you are assembling:
2. Slide the larger o-ring
over the connecting body
and into the groove in the
neck of the handpiece.
Slide the smaller o-ring
into the groove in the
metal connecting body.
23 kHz 36 kHz
Large o-ring black white
Small o-ring green blue
Caution
Do not put a nosecone of any type onto the handpiece before sterilization.
Surfaces covered by the nosecone may not be sterile.