Disconnect Suction Tubing, Irrigation Tubing, and the Handpiece
Shutting Down
the CUSA EXcel System
CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide 6-3
Disconnect Suction Tubing, Irrigation Tubing, and the Handpiece
1. Disconnect the green
L-shaped connector on the
manifold suction tubing
from the suction canister.
Then put the cap (if
available) on the canister lid.
2. Disconnect the green
L-shaped connector on the
contamination guard tubing
from the suction canister.
Then put the cap (if
available) on the canister lid.
Do not discard the
contamination guard.
3. Close the roller clamp on the I.V. tubing.
4. Disconnect the I.V. tubing from the irrigation tubing.
5. Open the irrigation
pump latch.