Integra Cusa Excel - Disconnect Suction Tubing, Irrigation Tubing, and the Handpiece; 1. Disconnect the green Lshaped connector on the manifold suction tubing from the suction canist...; 2. Disconnect the green Lshaped connector on the contamination guard tubing from the suction can...; 3. Close the roller clamp on the I.V. tubing.

Integra Cusa Excel
189 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Disconnect Suction Tubing, Irrigation Tubing, and the Handpiece
Shutting Down
the CUSA EXcel System
CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide 6-3
Disconnect Suction Tubing, Irrigation Tubing, and the Handpiece
1. Disconnect the green
L-shaped connector on the
manifold suction tubing
from the suction canister.
Then put the cap (if
available) on the canister lid.
2. Disconnect the green
L-shaped connector on the
contamination guard tubing
from the suction canister.
Then put the cap (if
available) on the canister lid.
Do not discard the
contamination guard.
3. Close the roller clamp on the I.V. tubing.
4. Disconnect the I.V. tubing from the irrigation tubing.
5. Open the irrigation
pump latch.