Clean the Tip Torquing Set
12-6 CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide
Clean the Tip Torquing Set
1. Using a soft cloth and mild cleansing solution or blood dissolving
detergent, manually clean the tip torquing base and the torque wrench.
2. Rinse the tip torquing base and torque wrench thoroughly with water.
3. Using a soft cloth, dry the tip torquing base and torque wrench.
Notice
Product damage will result if you do not follow these notices when cleaning the
Tip Torquing Set.
Do not use ultrasonic or automated washers.
Do not autoclave.
Do not use chlorinated substances such as bleach solution.
Do not clean with abrasives such as Ajax, Comet, or steel wool.