Clean the Sterilizer Case
Maintaining
the CUSA EXcel System
CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide 14-3
Clean the Sterilizer Case
Use a neutral detergent to clean the sterilizer case.
Maintain the Handpiece
Maintaining the handpiece involves lubricating the handpiece connector
o-rings or replacing the handpiece connector o-rings.
Lubricate the Handpiece Connector O-Rings
Keeping the o-rings lubricated prevents them from drying out and
cracking, which may cause handpiece cooling problems.
Frequency: Every month or every 80 hours of use, whichever comes first.
Items needed: A high viscosity lubricant such as silicone grease.
Items to be lubricated: The four o-rings on the handpiece connector water
prongs.
Replace the Handpiece Connector O-Rings
Regularly replacing the o-rings prevents excessive wear, which may result
in handpiece cooling problems.
Frequency: Every six months or 500 hours of use, whichever comes first.
Items needed: O-Ring maintenance kit, which includes o-rings, an
insertion tool, and instructions.
Items to be replaced: The four o-rings on the handpiece connector water
prongs.
Figure 14-1.
Replacing o-rings on the handpiece
connector water prongs
Notice
Do not clean the sterilizer case with abrasives. Product damage will result.
Insertion Tool

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals