EasyManuals Logo

Integra Cusa Excel User Guide

Integra Cusa Excel
189 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #181 background imageLoading...
Page #181 background image
Disassembly to Change Tips on the Sterile Field
S-10 Supplement: CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide
5. Connect the manifold
suction tubing and clip to
the suction port in the
nosecone.
6. Push one to three
manifold tubing clips
onto the handpiece cable.
Note: It is only necessary
to attach one to three
clips, but you can attach
all five clips in the
package.
7. Uncoil the CEM
cable and insert it
into the handpiece
and manifold tubing
clips.
Disassembly to Change Tips on the Sterile Field
Use this procedure only if you are disassembling the handpiece as part of
changing tips on the sterile field. If you are disassembling the handpiece
at the end of the surgical procedure, refer to Section 12, Disassembling
Handpieces in the CUSA EXcel System User’s Guide.

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals