Assembly
Supplement: CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide S-9
Attach the Manifold Tubing
If you receive the sterilized handpiece without the manifold tubing
already clipped onto the handpiece cable, you must attach the manifold
tubing now. Follow this procedure:
1. Remove the manifold tubing with clips from the packaging. Uncoil
some of the tubing.
2. Snap the handpiece clip
in place on the handpiece
housing.
3. Insert the irrigation tube
from the flue into the
handpiece clip.
4. Connect the manifold irrigation tubing to the Luer fitting on the
handpiece flue.