Environment
A-4 CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide
Environment
The system requires a minimum of one hour exposure at its operating
temperature range before you use it.
Standards and IEC Classifications
ATTENTION
Consult accompanying documents.
DANGER
Explosion risk if used with flammable anesthetics.
EQUIPOTENTIAL GROUND
Connect equipotential ground cable here.
POWER ON
Applies power at the AC Main Switch and the System Power Switch.
POWER OFF
Removes power at the AC Main Switch.
PROTECTIVE EARTH GROUND
CAUTION
To reduce the risk of electric shock, do not remove the cover. Refer
servicing to qualified service personnel.
Classified with respect to electrical shock, fire, mechanical, and other
specified hazards only in accordance with UL60601-1 and CAN/CSA
C22.2 No. 601.1.
Operating temperature
range
10°C (50°F) to 40°C (104°F)
Operating humidity range
15 to 80%, relative humidity, non-condensing
Operating atmospheric
pressure range
70 CPA to 106 CPA (10.15 to 15.38 psi)
Storage and shipping
temperature
-34°C (-29°F) to 65°C (149°F)
Storage and shipping
humidity
25 to 85% relative humidity, non-condensing