Reprocess the Sterilizable Torque Base
S-16 Supplement: CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide
Reprocess the Sterilizable Torque Base
This procedure was developed in accordance with the professional
recommendations of the Association for the Advancement of Medical
Instrumentation (AAMI) and the Association of periOperative Registered
Nurses (AORN). Cleaning efficacy has been validated using enzymatic
cleaning agents (for example, Klenzyme
®
and Enzol™) according to the
manufacturer's instructions.
After disassembling the tip from the handpiece, clean, then sterilize, the
sterilizable torque base. Refer to the following instructions.
Clean the Base
1. Remove all gross matter (blood, mucous, and tissue):
a. Dampen a soft cloth with a cleaning agent.
b. Thoroughly wipe all surfaces of the base. Follow the procedures
approved by your institution.
2. Soak at least two minutes in an enzymatic cleaning agent (for
example, Klenzyme
®
or Enzol™) according to the manufacturer's
instructions.
3. Scrub all surfaces with a small soft brush.
4. Rinse thoroughly with water.
5. Dry with a clean soft cloth.
Notice
Product damage will result if you do not follow these notices when cleaning the
handpiece:
Do not use ultrasonic or automated washers.
Do not use chlorinated substances such as bleach solution.
Do not clean the handpiece with abrasives such as Ajax, Comet, or steel wool.

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals