Items Needed
S-4 Supplement: CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide
Figure 4.
Sterilizable tip torquing base
You can use the sterile torque wrench to change as many as five tips in one
surgical procedure. Do not use the sterile torque wrench in more than one
surgical procedure. Do not sterilize the sterile torque wrench.
Figure 5.
Sterile torque wrench (23 kHz or
36 kHz)
Manifold tubing is sterile, single use only. You can resterilize unused
manifold tubing one time. You have two options for attaching the tubing
to the handpiece:
Attach the manifold tubing on the sterile field.
Attach the manifold tubing before sterilizing the handpiece.
Figure 6.
Sterile manifold tubing
23 kHz
(Green)
36 kHz
(Blue)