Sterilize the Handpiece
10-4 CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide
With Manifold tubing Attached – Cycles that include the CUSA EXcel Manifold
Tubing contained five BIs in each of three CUSA EXcel handpiece
assemblies for a total of 15 BIs per run. The locations were the following:
In the center of the manifold suction tubing (spore strip or dot)
In the center of the manifold irrigation tubing (spore strip or dot)
In the center of an unattached tip (spore suture)
In the center of an unattached flue tubing with tip cleaner inserted into
the flue (spore suture)
Under the front o-ring of the handpiece (spore suture).
Acceptance Criteria
In order for the CUSA EXcel handpiece to be considered validated, none
of the biological indicators demonstrated growth of the indicator
organism. A 6-log reduction at the half cycle exposure time indicates that
a full cycle (twice the half cycle exposure time) will produce sufficient
lethality to effect at least a 12-log reduction and provide a 10
-6
sterility
assurance level (SAL).
Packaging
Wrapped: Sterilizer case double wrapped in hospital CSR material.
Flash: Sterilizer case unwrapped.

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals