EasyManuals Logo

IRIS iQ 200 Sprint User Manual

IRIS iQ 200 Sprint
215 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
Iris Diagnostics, a Division of Iris International, Inc.
iQ
®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 6-3
6. Troubleshooting
Troubleshooting Guide
System locked up
System fails to start measuring or takes excessive time to start
measuring.
Instrument is waiting for chemistry result (2min wait time)
System is offline
Low or empty Lamina
1. Check Lamina reservoir
2. Check for kinked Lamina tubing
3. Check if the fill pump is running
Peristaltic pump housing interlock
1. Check cassette sensor
2. Check locking screw sensor
Air system not pressurizing
1. Leak in air system
2. Check if the air pump is running
Auto focus failure
Camera cable is not connected to the RP computer
Could not capture background
1. Field of view of flow cell is not complete. Perform collimation.
2. Light level incorrectly set due to high light level set size per hex.
value
3. Strobe bulb failure
Flow errors
1. Run control rack
2. Verify proper flow using Runfile analyzer
3. Possible fluidic obstruction or loose fitting within fluidic path
Focus is too far out for system to find starting point
1. Perform manual focus in Service mode
Failed OBA nudger
1. Verify that nudger operates in Service mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IRIS iQ 200 Sprint and is the answer not in the manual?

IRIS iQ 200 Sprint Specifications

General IconGeneral
BrandIRIS
ModeliQ 200 Sprint
CategoryLaboratory Equipment
LanguageEnglish