Iris Diagnostics, a Division of Iris International, Inc.
iQ
®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 3-28
3. Components
SPA SIC Push-Button Switches
There is a single push-button on the SPA SIC that manually resets the
Hard Crash condition if the incoming Hard Crash signal (from the Hard
Crash switch on the SPA arm) is no longer asserted.
Connect. Type Connects to Function
J1 34-pin ribbon header Backplane J19 Main signals to SMCS
J2 12-pin female
SPA Assy Tether
ESN
SPA Assy ESN
J3 12-pin female ESN
SPA Vertical Motor
ESN
SPA Vertical Motor ESN, DC
motor and encoder
J4 12 pin female ESN
SPA Rotation
Motor
SPA Rotation Motor ESN, DC
motor and encoder
J5 12-pin female
SPA Crash
Detector Switches
SPA Crash Detection
J6 12-pin female
SPA Bottom
Sensor switch ESN
SPA Bottom Sensor ESN and
switch
J10 10-pin ribbon header SPA Opto Bd
SPA vertical & rotational
homes and other limit.