EasyManua.ls Logo

IRIS iQ 200 Sprint

IRIS iQ 200 Sprint
215 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Iris Diagnostics, a Division of International Remote Imaging Systems, Inc.
iQ
®
200 Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 4-3
4. Adjustments
Accessing the Service Screen
Press the Windows key on the keyboard (located between Ctrl and Alt
keys). The Windows toolbar will be displayed at the bottom of the screen.
On the tool bar, left click on SM. The screen will display the SM screen.
Select FileSwitch to Service Application.
Select Yes to switch to Service Application. The system will prompt to
switch off the Microscopy Module. When connections are established
with the Results Processor, the system will prompt you to turn the
Microscopy Module back on.
After about 1 minute, the Service Screen will be displayed.

Table of Contents