200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 9-13
9. Preventive Maintenance
Auto-Focus run
1. Perform an Auto-Focus by placing a tube of Focus Solution in the
Control rack in position 5. Verify Auto-Focus completes
successfully.
Performance verification
Calibration run
Run Calibration and verify REF value is between 1.4 – 2.7
Control run
Run 10 Positive/Negative Controls, verify the Positive Control runs are
within range with a CV of <5% using Runfile Analyzer.
Reporting
Report all the results on the PM check list form and send the form to your
regional service contact at Iris Diagnostics.