Iris Diagnostics, a Division of Iris International, Inc.
iQ
®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 6-7
6. Troubleshooting
Waste well overflow
1. Waste pump failure
2. Float sensor failure in Lamina tank (Overflow from lamina tank)
3. Waste well overflow detector is wet
4. Overflow detector failure
Event Log Error Messages
Error Message in
Maintenance
SPA Failure
mode
FBA Failure
mode
OBA Failure
mode
SPA State 102 or 121,
Sequence 27 or 29, Error
timeout
Vertical up
position fails to
reach down
MC State 153 Sequence 6 Focus dial not
moving