200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 6-8
6. Troubleshooting
Error Message in
Maintenance
SPA Failure
mode
FBA Failure
mode
OBA Failure
mode
SPA State 0 Sequence 18
Error: low-level controller,
Detail: scanner error. Waste
well overflow.
Lamina sensor
in lamina tank is
stuck down
position - lamina
overfills into
waste well
SPA State 3 Sequence 1
Error: timeout
Spa vertical
motor failed to
home
SPA State 3 Sequence 1
Error: timeout
EP is not
homing
SPA State 3, Sequence 1
Error: timeout, Detail: N/A
EP not reaching
to position
during auto
focus. This
causes auto
focus to fail, but
after rack is
ejected and EP
attempts to go
home during PH.
SPA State 4 Sequence 1
Error: timeout
Spa rotational
motor failed to
home
SPA State 8 Sequence 1
Error: timeout
Air system
unable to reach
setpoint
SPA State 99 Sequence 26
Error: Timeout Detail: NA
EP not reaching
to position
SPA State 116, 135
Sequence 27 or 29 Error:
timeout
Air system
unable to reach
setpoint
SPA State 42 Sequence 13
Error: Low-level controller.
Detail: scanner error. Spare.
EP not reaching
to position
SPA State 90 Sequence 25
Error: timeout
Spa rotational
motor failed to
home