Iris Diagnostics, a Division of Iris International, Inc.
iQ
®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 9-6
9. Preventive Maintenance
4. Remove and replace the cassette on the Lamina Pump (PN#700-
3822). Unscrew the connectors to release the cassette tubing.
Once the new cassette has been secured, remove the hemostat.
5. Verify that all the peristaltic pump doors are closed properly.