EasyManua.ls Logo

Keysight Technologies 8163A - Page 122

Keysight Technologies 8163A
466 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8163A/B, 8164A/B, 8166A/B User’s Guide 121
Power Measurement 3
You can generate input triggers in any of the following ways:
trigger the Input Trigger Connector on the rear panel of your
instrument,
set <Trigger Configuration> to <Loopback> so that an output trigger
automatically generates an input trigger, or
use the
:TRIGger
GPIB command, see your mainframe’s Programming
Guide.
How to Use Output Triggering
You can configure 8163A/B Series Power Meter modules to output a
trigger when the instrument performs certain tasks.
To set your module’s Output Trigger Configuration:
1See How to Set the Trigger Configuration on page 68 for how to
configure the trigger connectors.
2 Move to the Power Meter channel and press the [Menu] softkey.
3Move to <Output Trigger Mode>, by using the cursor key, and press
Enter. You will see the screen in Figure 71.
Figure 71 Output Trigger Mode
4 Move to one of the following, by using the cursor key:

Table of Contents

Related product manuals