8163A/B, 8164A/B, 8166A/B User’s Guide 121
Power Measurement 3
You can generate input triggers in any of the following ways:
trigger the Input Trigger Connector on the rear panel of your
instrument,
set <Trigger Configuration> to <Loopback> so that an output trigger
automatically generates an input trigger, or
use the
:TRIGger
GPIB command, see your mainframe’s Programming
Guide.
How to Use Output Triggering
You can configure 8163A/B Series Power Meter modules to output a
trigger when the instrument performs certain tasks.
To set your module’s Output Trigger Configuration:
1See How to Set the Trigger Configuration on page 68 for how to
configure the trigger connectors.
2 Move to the Power Meter channel and press the [Menu] softkey.
3Move to <Output Trigger Mode>, by using the cursor key, and press
Enter. You will see the screen in Figure 71.
Figure 71 Output Trigger Mode
4 Move to one of the following, by using the cursor key: