380 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
Error
One or more of the following reasons may cause this error message:
error message Offset table is empty
description The user has tried to activate the Offset(Lambda) function, but the table of wavelength dependent offsets is
empty.
affected modules All attenuators
action Please construct a table with at least one entry, before activating this mode.
error message Lambda offset active
description The user has tried to program the table of wavelength dependent offsets, but the Offset(Lambda) mode is
activated.
affected modules All attenuators
action Please deactivate the Offset(Lambda) mode first.
firmware revision all
modus stepped sweep
parameter(s) lambda logging on
action switch lambda logging off