400 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
Power Meters & Interface Modules with Optical Head
error message E10010: Invalid Caldata
description The Caldata loaded on the module is invalid. Either the caldata version number is not
compatible with the firmware or the caldata is corrupt (checksum incorrect)
test value (no test value)
action Please return the unit to your local service center for repair
error message E10012: Dual ported RAM write/read test failed
description Writing bit patterns and reading them back to/from the dual ported ram failed
test value (no test value)
action Please return the unit to your local service center for repair
error message E10013: FPGA not correctly initialized
description Loading the FPGA with data failed. This indicates a defective FPGA or an FPGA access
problem
test value (no test value)
action Please return the unit to your local service center for repair
error message E1x020: Communication with head failed
description An external head is available but it was not possible to access the registers of the
head’s FPGA
test value x refers to the module channel tested
action Please return the unit to your local service center for repair