EasyManuals Logo

Keysight Technologies 8163A User Manual

Keysight Technologies 8163A
466 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #419 background imageLoading...
Page #419 background image
418 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
Optical Switch modules
error message InitError() failed!
description Initialization of internal error buffer failed.
test value
action Please return the unit to your local service center for repair
error message Invalid Caldata!
description The calibration data loaded on the module is invalid.
Either the version number is nir compatible with the firmware or the calibration data is
corrupt (checksum incorrect).
test value
action Please return the unit to your local service center for repair
error message InitModule() failed!
description Module initialization failed.
note Normally something more descriptive appears (i.e. one of the other messages mentioned
here.)
test value
action Please return the unit to your local service center for repair
error message SanityChecks():channel X, configArray != -1
SanityChecks():channel X, configs == 0
SanityChecks() failed
description Calibration data seems not to be correct
test value X Channel number
action Please return the unit to your local service center for repair

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 8163A and is the answer not in the manual?

Keysight Technologies 8163A Specifications

General IconGeneral
BrandKeysight Technologies
Model8163A
CategoryMultimeter
LanguageEnglish

Related product manuals