418 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
Optical Switch modules
error message InitError() failed!
description Initialization of internal error buffer failed.
test value
action Please return the unit to your local service center for repair
error message Invalid Caldata!
description The calibration data loaded on the module is invalid.
Either the version number is nir compatible with the firmware or the calibration data is
corrupt (checksum incorrect).
test value
action Please return the unit to your local service center for repair
error message InitModule() failed!
description Module initialization failed.
note Normally something more descriptive appears (i.e. one of the other messages mentioned
here.)
test value
action Please return the unit to your local service center for repair
error message SanityChecks():channel X, configArray != -1
SanityChecks():channel X, configs == 0
SanityChecks() failed
description Calibration data seems not to be correct
test value X Channel number
action Please return the unit to your local service center for repair