8163A/B, 8164A/B, 8166A/B User’s Guide 417
Error Messages 14
The following error codes are attenuator module specific:
Error Creating message queue!
•Error Init MCB!
Error spawning tLowlvlMsgHandler!
Error spawning tKeyPressed!
Error spawning tTempCtrl!
Error spawning tPowCtrl!
Error checking firmware CRC!
b shows the measured values for an amplification by 10
c shows the measured values for an amplification by 100
action Please return the unit to your local service center for repair
error message E10063: Lambda Dac adj failed (a, b, c, d)
description Verifies the gain of the responsiveness correction stage
test value a if 1, the quotient d could not be calculated, otherwise 0
b shows the difference between DAC settings 1024 and 2048
c shows the difference between DAC settings 2048 and 4096
d shows the quotient of b divided by c
action Please return the unit to your local service center for repair