8163A/B, 8164A/B, 8166A/B User’s Guide 377
Error Messages 14
Errors appearing on pop-up menus
Error on Module
error message Calibration data corrupted
description The calibration data loaded on the module is invalid. Either the caldata version number is not compatible
with the firmware or the caldata is corrupt (checksum incorrect)
affected modules all modules
action Please return the unit to your local service center for repair
error message Laser safety error
description The laser control circuit has switched of the laser. Either the laser current or the laser power was too high.
affected modules TLS
action Try to run the automatic realign from the menu or to operate the laser at a lower power level. If the failure
persists, please return the unit to your local service center for repair.
error message Temperature out of range
description The temperature to operate the tunable laser source module was measured outside given limits.
affected modules TLS
action Verify the fan is working and the ventilation system is not blocked. If the failure persists, please return the
unit to your local service center for repair.