122 8163A/B, 8164A/B, 8166A/B User’s Guide
3 Power Measurement
• <Averaging Time Over>, a trigger is output when the averaging time
period of a measurement finishes.
• <Meas Started>, a trigger is output when the averaging time period
of a measurement begins.
• <Disabled>, the output trigger mode is disabled.
5Press Enter.
Dual Power Meters - Master and Slave Channels
For the Keysight 81635A Dual Power Sensor and the Keysight 81619A
Dual Optical Head Interface module, channel 1 is the master channel and
channel 2 is the slave channel. The master and slave channels share the
same software and hardware triggering system. For some commands,
setting parameters for the master channel sets the parameters for the
slave channel. In these cases, you may only set parameters for the slave
channel by setting master channel parameters.
The parameters listed in Table 12 on page -122 can only be set or viewed
using the master channel.
Table 12 Parameters that can only be set using the master channel
Parameter
<Zero>
<MinMax mode>
<Reset MinMax>
<Data points>
<Averaging time>
<Range mode>
<Input trigger mode>
<Output trigger mode>