382 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
Tunable Laser Sources
Initialization Tests
These tests are executed every time the instrument is powered on, before
the selftests start.
error message Initialization Error (AttInitialize(), error value)
description Error during attenuator initialization. No ready message from attenuator DSP after ...
error value 1 reset
2 init motor
4 find index
8 move to min position
action Please return the unit to your local service center for repair
error message Initialization Error (AttMovePos(zeroPos), error value)
description No ready message from attenuator DSP after moving to zero position taken from caldata
error value 8 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (CalData CRC wrong: crc1, stored: crc2!)
description The checksum stored in the caldata does not match the calculated checksum
error value crc1 calculated checksum
crc2 checksum read from caldata
action Please return the unit to your local service center for repair