8163A/B, 8164A/B, 8166A/B User’s Guide 445
Cleaning Instructions 15
For Class 3B lasers
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the interface by rubbing a new, dry cotton-swab over the surface
using a small circular movement.
2 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure particularly when there is greasy dirt on the interface:
1 Moisten a new cotton-swab with isopropyl alcohol.
2 Clean the interface by rubbing the cotton-swab over the surface using
a small circular movement.
3 Take a new, dry soft-tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular
movement.
4 Blow away any remaining lint with compressed air.
WARNING
The built-in laser diode is active whenever the instrument is powered on,
therefore disabling the output is not sufficient to establish eye and skin
safe conditions.
CAUTION
Protect your extra-high power tunable laser from damage:
Clean the instrument’s output connector and all interfaces of attached
fibers and termination devices to avoid burning-in dust and other
residue!
Do not leave plastic caps on the output connector or on the end of a
connected fiber when you activate the laser – you risk damaging the
connector surface with deposit from hot plastic!
Make sure the optical path is terminated properly and confirm that the
termination can cope with an optical power level of over +18.5 dBm or
70 mW!
Use a metal cap to cover the laser outputs when open. Terminate
open patchcord ends with a commercially available “fiber optic light
trap”.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 8163A and is the answer not in the manual?

Keysight Technologies 8163A Specifications

General IconGeneral
BrandKeysight Technologies
Model8163A
CategoryMultimeter
LanguageEnglish

Related product manuals