294 8163A/B, 8164A/B, 8166A/B User’s Guide
10 Applications
If there is another Power Meter channel to reference, you will be prompted
to perform this reference measurement from the user interface by a
prompt box similar to the box in Figure 180 on page -293. Restart this
procedure at
step 20.
If you are viewing the PACT Reference Measurement Finished box,
reference sweeps have been performed for all selected Power Meters.
Figure 181 PACT Reference Measurement Finished Box
How to Perform a Loss Measurement
23 To return to the Setup screen, pressing [Menu], move to <Setup>, and
press Enter.
NOTE
If you change any measurement parameters, the reference measurement
becomes invalid.