412 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
DFB Laser Sources
error message E10010: Invalid Caldata
description The Caldata loaded on the module is invalid. Either the caldata version number is not
compatible with the firmware or the caldata is corrupt (checksum incorrect)
test value (no test value)
action Please return the unit to your local service center for repair
error message E10012: Dual ported RAM write/read test failed
description Writing bit patterns and reading them back to/from the dual ported ram failed
test value (no test value)
action Please return the unit to your local service center for repair
error message E10013: FPGA not correctly initialized
description Loading the FPGA with data failed. This indicates a defective FPGA or an FPGA access
problem
test value (no test value)
action Please return the unit to your local service center for repair
error message E10040: 10bit ADC test failed (a, b, c)
description The 10bit ADC has returned a value other than 512
test value a shows the ADC reading
b shows the lower limit
c shows the upper limit
action Please return the unit to your local service center for repair