EasyManua.ls Logo

Keysight Technologies 8163A - Laser Safety Information

Keysight Technologies 8163A
466 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8163A/B, 8164A/B, 8166A/B User’s Guide 23
Getting Started 1
Laser Safety Information
The laser sources specified by this user guide are classified according to
IEC 60825-1.
The laser sources comply with 21 CFR 1040.10 except for deviations
pursuant to Laser Notice No. 50 dated 2007-June-24.
Table 1 Initial Safety Information for Laser Source Modules
81650A 81651A 81652A 81654A
Laser type FP-Laser
InGaAsP
FP-Laser
InGaAsP
FP-Laser
InGaAsP
FP-Laser
InGaAsP
Wavelength 15nm) 1310 nm 1550 nm 1550/1625 nm 1310/1550 nm
Max. CW output power
1
<1.8 mW <1.8 mW <1.8 mW <1.8 mW
Beam waist diameter 9 µm 9 µm 9 µm 9 µm
Numerical aperture 0.1 0.1 0.1 0.1
Laser Class according to
IEC 60825-1 (2007)
1111
Max. permissible CW output power 15.6 mW 10 mW 10 mW 15.6 mW/10 mW
1
Max. CW output power is defined as the highest possible optical power that the laser source can produce at its output connector.

Table of Contents