Page #162 background imageLoading...
Page #162 background image
8163A/B, 8164A/B, 8166A/B User’s Guide 161
Tunable Lasers 5
3 To pass the master trigger to the slaves, set up the mainframe through
the “Config” button under the screen:
Config -> Trigger -> Feedback
Note that the master laser must always be turned on, if one or more slaves
are on.
How to Increase Linewidth
You can use coherence control to increase the linewidth of the optical
signal output from your Tunable Laser module.
Enabling the coherence control increases the linewidth of the optical
output signal to between 50 and 500 MHz (typically). Coherence control
greatly reduces interference effects and therefore improves the power
stability in sensitive test setups.
Some models also allow adjustment of the linewidth broadening with the
"Coherence level" setting. A level of 100% corresponds to minimal
broadening. The default level is usually a good setting.
To enable coherence control:
1 Move to the Tunable Laser channel and press [Details].
2Move to [Mod Src] and press E
NTER.
3Move to <Coherence Ctrl.>, Coherence Control, to increase linewidth
to approximately 500 MHz.
4Press Enter. The text CC appears in the Tunable Laser channel.
How to Set the Output Power of a Modulated Signal
All external modulation modes require you to set the output power. To set
the output power, see How to Set the Output Power of a Modulated Signal
on page 155.
How to Configure the Modulation Output
If your Tunable Laser module has a BNC output connector on its front
panel, to synchronize your external measuring equipment to the
modulation of the module:
1 Move to the Tunable Laser channel and press Menu.
How to Set the BNC Output Line Mode as a Modulation Output
2Move to <BNC Output> and press [OK].
3Move to <Modulation>, by using the cursor key. A TTL-level signal is
output, if the optical output is digitally modulated, that is, if you
choose <Internal>, <External Digital>, or <Backplane> as the chosen

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 8163A and is the answer not in the manual?

Keysight Technologies 8163A Specifications

General IconGeneral
BrandKeysight Technologies
Model8163A
CategoryMultimeter
LanguageEnglish

Related product manuals