8163A/B, 8164A/B, 8166A/B User’s Guide 381
Error Messages 14
firmware revision <= 2.5x
modus continuous sweep
parameter(s) stop wavelength <= start wavelength
sweep time too small (< 0.001s)
sweep time too large (t > 400s)
trigger frequency too high (f > 10kHz)
lambda logging:
- coherence control active
- cycles ≠ 1
- trigger ≠ step finished
action check settings and correct where appropriate
firmware revision > 2.6
modus continuous sweep
parameter(s) stop wavelength <= start wavelength
sweep time too small (< 0.001s)
sweep time too large (t > 400s)
trigger frequency too high (f > 40kHz)
lambda logging:
- coherence control active
- trigger ≠ step finished
action check settings and correct where appropriate