8163A/B, 8164A/B, 8166A/B User’s Guide 399
Error Messages 14
error message E10080: Laser <a> control for safety test failed
description To test the current safety comparator the laser current needs to be set to two different
values
test value a laser number
action Please return the unit to your local service center for repair
error message E10081: Safety comparator <a> test failed (b, c)
description Checks whether laser current safety comparator includes at the right conditions a safety
interrupt at the CPU
test value a laser number
b shows whether the interrupt flag was raised when the current was lower than
the limit
c shows whether the interrupt flag was raised when the current was higher
than the limit
action Please return the unit to your local service center for repair