EasyManuals Logo

Keysight Technologies 8163A User Manual

Keysight Technologies 8163A
466 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #160 background imageLoading...
Page #160 background image
8163A/B, 8164A/B, 8166A/B User’s Guide 159
Tunable Lasers 5
Figure 92 Wavelength Locking
To enable wavelength locking:
1 Move to the Tunable Laser channel and press [Details].
2Move to <Mod Src> and press Enter.
3Move to <Wavel. Locking>, by using the cursor key, and press Enter.
The text λLock appears in the Tunable Laser channel.
External Digital Modulation using Input Trigger Connector
External digital modulation uses a TTL-level signal. Apply this signal to the
Input Trigger connector on the rear panel of your mainframe. For
information on external digital modulation, see External Digital Modulation
on page 156.
l
V
+5 V
-5 V
CAUTION
A maximum of 5 V can be applied as an external voltage to the Input
Trigger connector.
Take care not to use the Trigger Output connector or the Remote
Interlock connector for modulation. Do not apply an external voltage to
these connectors.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 8163A and is the answer not in the manual?

Keysight Technologies 8163A Specifications

General IconGeneral
BrandKeysight Technologies
Model8163A
CategoryMultimeter
LanguageEnglish

Related product manuals